Continuous Particulate Measurement

Continuous Particulate Measurement



Order

Description

Sensors’ Continuous Particulate Measurement utilizes an ion mobility technique (based on the Pegasor® PPS-M) whereby clean air is ionized and mixed with the sample, charging the particles. As charged particles exit the sensor (Faraday cup) their charge is measured with a sensitive electrometer. Accordingly, the SEMTECH CPM is a real-time PM measurement system that can be configured to measure either PM Mass or PM Number concentrations. The module can be employed as a standalone analyzer, or integrated with the SEMTECH MPS and PFS to provide a complete SEMTECH PM measurement solution for both real-world or laboratory applications.


Feature

  • 1065 and EU ECE Reg. 49 compliant: The SEMTECH CPM meets the preliminary requirements for US EPA Part 1065 and EU ECE Reg. 49 compliance requirements for in-use testing, when integrated with SEMTECH MPS and PFS.
  • PM Mass or Number: The module may be calibrated to accurately measure either particle mass or number.
  • Flow-through transducer design: The CPM is designed to extract a diluted aerosol sample from either a third party diluter, the MPS or a CVS sampling system. This results in improved performance (humidity effects, pulsation effects, etc.) and greatly reduces service intervals.
  • Protection from contamination while not in operation: The CPM module is configured with a zero air solenoid to enable real-time zeroing as well as providing a stand-by feature.
  • Power supply monitoring: Power can be either 12 VDC, 110 VAC or 220 VAC, with both current and voltage monitoring.
  • Rugged construction: This unit can be used off-road and in other harsh environments.
  • Shock Resistance: The mechanical design has been optimized for resistance to shock and vibration, ensuring data acquisition in the most harsh environments.
  • Design Details: Over a decade of experience in in-use emissions testing has gone into the design details of the SEMTECH ECOSTAR system, including:
    • EMI protection, including gaskets, filters
    • Stress relief for pneumatic connections
    • Channels for cable management
    • Standard Swagelok™ bulkhead connectors
    • Rugged Deutsch connectors for power and auxiliary connectors
    • Handles that lock to other SEMTECH ECOSTAR modules for stable system integration

Spesification

Measurement Technique

Ion Mobility

Minimum Particle Size

> 23nm (dependent on selected trap voltage)

Maximum Particle Size

< 2.5μm (dependent on measured particle size distribution)

Concentration

From 1μg/m3 to 250 mg/m3

Response Time

< 0.3 sec

Maximum Data Rate

10Hz (high speed internal averaging)

Power Requirement

12 VDC; 110-220 VAC

Storage Temperature

Dry -10°C to 60°C

Operating Temperature

-10°C to 45°C ambient

Communications

USB

Dimensions (WxDxH)

  1. 17.2” x 12.2” x 7.1"

Weight:

20kg 44lbs

Electromagnetic Interference & Susceptibility

CE Standards: ICE 61326:2002-2