TIME2600 Coating Thickness Gauge

TIME2600 Coating Thickness Gauge



Order

Description

Coating Thickness Gauge TIME®2600


Feature

Features

  • Two operation principles are adapted: magnetic induction (ferrous) and eddy current (non-ferrous) to take non-destructive measurements
  • 6 types of probes are available for different applications
  •  Features two working modes: DIRECT and BATCH& two measuring ways: CONTINUE and SINGLE
  • Statistics include the mean, maximum, minimum, test numbers and standard deviation.
  •  Memory of 640 data
  • Two calibration methods for better correction
  • Integrated with printer to print the statists values if needed
  • Low battery indication and error alarm
  • Backlight for the screen
  • Auto or manual shutdown
  • Conform to the standards of DIN, ISO, ASTMBS.

Spesification

Technical Specification

Probe types

F

N

Measuring methods

magnetic induction

eddy current

Measuring range

0 ~1250 μm

0 ~1250 μm, 0 to 40μm (for chrome plate on copper)

Minimum resolution

0.1μm

Tolerance

Zero point calibration

±(3%H+1)μm

± (3%H+1.5)μm

H means the thickness of tested piece

Two points calibration

±[(1~3)%H+1]μm

±[(1~3)%H+1.5]μm

H means the thickness of tested piece

Measuring condition

Min. curvature radius (mm)

Convexity 1.5

Convexity 3

Min. testing area diameter (mm)

Ø7

Ø5

Critical thickness of substrate(mm)

0.5

0.3

Standards

DIN,ISO,ASTM,BS

Calibration

Zero and foil calibration

Interface

RS232

Statistic

Number of measurement, mean, standarddeviation, maximum and minimum

Data memory

640 readings

Limits

Adjustable with alarm

Power

NiMH rechargeable battery

Operating environment

Temperature: 0~40?

Humidity: 20%~90%

No strong magnetic field

Dimensions (mm)

230×86×47